C3050

C3050/C3051 is an instrument capable of measuring all parameters according to Vincent’s method. All parameters can be measured
at once and shown on the wide display. C3051 version has an Ethernet connection instead of USB.
This instrument is built with the same quality as the instruments in the whole C3000 series. As such, it has the same features as the
other versions. All 3 gold plated BNC connectors each accept different electrodes of which 2 connectors also accept conductivity
electrodes. There is no interference between the channels when measuring pH/ORP/Ion and conductivity electrodes in the same
solution.
Besides the pre-programmed pH buffers and EC standards, you can also add your own buffer and standard tables. Not just a certain
value at a certain temperature but the complete temperature related table of your specific buffer. Add up to 5 pH buffers and 3 EC
standards of your own choice and use them as if they are built-in.
Two temperature inputs are independent from measurement channels. ATC for each channel can be selected from any of the
temperature inputs. The device can be connected to a PC and completely controlled via either USB/RS232 or Ethernet/RS232.
Both software and communication protocol can be downloaded from our website.

Category:

Description

Multichannel up to 3 measurements can be performed at the same time and simultaneously or individually shown on the display. All parameters according to Vincent’s method can be shown on the display. Each channel can measure 2 times per second.

No interference between pH/ORP and conductivity electrodes in the same solution

Custom calibration tables allow the user to add complete buffer/standard-temperature relation tables. With this feature the built-in tables can be extended with your own tables. Tables can be entered via a device menu or uploaded from a PC.

Direct rH 2 and μW measurement when using a pH, ORP and EC electrode. All 5 parameters can be displayed simultaneously.

Wide Display shows all channels at the same time including temperature and date/time.

Stability indicator ensures visualisation when the measurement has stabilised.

Stability algorithm ensures stable readings with ability to detect fast changes.

Hold function allows to freeze the display

Selectable resolution for more stable readings on mV and pH.

Range lock for conductivity measurements.

Capacitive compensation eliminates the capacitive component of the electrode and cable when measuring low conductivities.

Galvanic isolated USB interface eliminates ground loop effects when connected to a PC (C3050 only).

Ethernet connection allows remote access to the device via LAN or Internet (C3051 only)

Pre-programmed standards
pH: 1.68, 2.00, 4.00, 4.01, 6.87, 6.99, 9.18, 9.21, 10.01, 12.00, 12.45 (at 25°C)
Conductivity: 1413 μS/cm, 12.88 mS/cm, 111.8 mS/cm (at 25°C)

Free software and firmware updates downloadable from www.consort.be